TY - CHAP U1 - Konferenzveröffentlichung A1 - Knopf, Christoph A1 - Klöckner, Stephan A1 - Kanoun, Olfa A1 - Himmel, Jörg T1 - Experimental visualisation of potential distribution in electrosurgery T2 - 10th International Multi-Conference on Systems, Signals & Devices (SSD) N2 - The transurethral resection (TUR) is a standard technique in urological treatment procedures. Both, monopolar and bipolar electrosurgical systems, are used for TUR. Whereas electrical and physical processes in surgery surroundings are well understood for monopolar systems, there is no sufficient data base for the assessment of the processes with the use of bipolar systems. In this context a multi-electrode measuring system was developed to visualize the spatial potential distribution around bipolar electrosurgical devices as a first step to risk analysis. To simulate the anatomic surroundings of a transurethral surgery a cylinder filled with isotonic saline solution was used as a complexity reduced experimental environment. Y1 - 2013 UR - https://ieeexplore.ieee.org/document/6564156 SN - 978-1-4673-6457-7 SB - 978-1-4673-6457-7 U6 - https://doi.org/10.1109/SSD.2013.6564156 DO - https://doi.org/10.1109/SSD.2013.6564156 N1 - 2013 10th International Multi-Conference on Systems, Signals & Devices (SSD) Hammamet, Tunisia, March 18-21, 2013 SP - 1 EP - 5 CY - Hammamet, Tunisia ER -