Experimental visualisation of potential distribution in electrosurgery
- The transurethral resection (TUR) is a standard technique in urological treatment procedures. Both, monopolar and bipolar electrosurgical systems, are used for TUR. Whereas electrical and physical processes in surgery surroundings are well understood for monopolar systems, there is no sufficient data base for the assessment of the processes with the use of bipolar systems. In this context a multi-electrode measuring system was developed to visualize the spatial potential distribution around bipolar electrosurgical devices as a first step to risk analysis. To simulate the anatomic surroundings of a transurethral surgery a cylinder filled with isotonic saline solution was used as a complexity reduced experimental environment.
Author: | Christoph Knopf, Stephan Klöckner, Olfa Kanoun, Jörg Himmel |
---|---|
URL: | https://ieeexplore.ieee.org/document/6564156 |
DOI: | https://doi.org/10.1109/SSD.2013.6564156 |
ISBN: | 978-1-4673-6457-7 |
Parent Title (English): | 10th International Multi-Conference on Systems, Signals & Devices (SSD) |
Place of publication: | Hammamet, Tunisia |
Document Type: | Conference Proceeding |
Language: | English |
Year of Completion: | 2013 |
Creating Corporation: | IEEE |
Release Date: | 2019/04/15 |
First Page: | 1 |
Last Page: | 5 |
Note: | 2013 10th International Multi-Conference on Systems, Signals & Devices (SSD) Hammamet, Tunisia, March 18-21, 2013 |
Institutes: | Fachbereich 4 - Institut Mess- und Senstortechnik |
DDC class: | 600 Technik, Medizin, angewandte Wissenschaften / 621.3 Elektrotechnik, Elektronik |
Licence (German): | ![]() |